Cleaning solution for "Laktan 1-4"

    For cleaning of the measuring channel of the instruments "Laktan 1-4" we devised the set of reagents for removal of remains of protein and milk stone. High efficiency of these detergents was proved by prolonged trials at our enterprise..

 

clean[1].jpg

The milk quality analyzer "Laktan 1-4" is a complicated device, which requires a careful treatment and maintenance in the process of operation and transportation. To provide normal operation of the device and receive accurate and precise measurements results it is required to carry out daily cleaning of the analyzer's measuring cell from the residuals of fat, protein and milk stone, which accumulate and distort the measurements results that can subsequently lead to the breakage of the instrument.
     Cleaning of the measuring cell with domestic detergents (dishwashing detergents or laundry detergents) is not effective due to bleach powdering, dye stuffs, organic compounds and aromatizers frequently contained in domestic detergents which can accumulate in the measuring cell and add to additional contamination of the device. This is why we devised a special technique of cleaning of the measuring cell and removal of fat, protein and milk stone which ensures its cleanness and sterility and guarantees reliable and no-failure operation of the device. We recommend using Reagent N1 and Reagent N 2 for daily cleanings of the analyzer. These chemicals are non-toxic and have no harmful effect. Water solutions of these detergents can be prepared once a month and used during the whole month.

NOTICE: In case of failure or breakage of the analyzer caused by contamination of the measuring channel or improper cleaning of the device the manufacturer reserves the right to withhold the manufacturer's warranty.

 

*General Imformation

INSTRUCTION


How to use reagent #1 and reagent #2 to remove the remains of protein and milk stone


1.Pour a clean tap water, heated to 40C, in the cup, clean the device.
2.Heat the tap water to 50C, dissolve reagent # 1 (1g on 100ml water), clean the device, changing cleaning solution three times.
3.Clean the device with the clean tap water until the water becomes clean.


When the device has been used for one week, it is necessary to clean it with reagent #2.


1.Heat the tap water to 50C, dissolve the reagent # 2 (0,5g for 100ml water)
2.After cleaning the device with reagent #1 and water, clean it with reagent #2, changing the cleaning solution 3 times.
3.Clean the device with the clean tap water until the water becomes clean